# LRCD220
As the LRCD210 with the addition of 2 wire fast testing, plus phase to phase and phase to earth measurement.
Both instruments will not trip RCDs rated at 30 mA or greater during loop testing, even electronic RCDs.
Voltage Measurements | 0 V – 500 V |
Voltage Accuracy | ±2% ±2 digits |
Frequency Measurement Range | 25Hz to 450Hz |
Frequency Accuracy | 25.0Hz to 199.9Hz ±0.1Hz 200Hz to 450Hz ±1Hz |
Phase Rotation Indicator (LRCD220 only) | Three wire identification of phase rotation. |
Frequency | 45Hz to 65Hz |
Loop Testing | Loop ranges (to EN 61557-3) |
No-Trip Loop Test (3 Wire, Line To Earth) | LRCD200/210 200V – 280 V 45Hz to 65Hz LRCD220 100V – 280 V 45Hz to 65Hz |
Nominal Test Current | 15 mA |
Loop Accuracy | 0.01 ½ – 9.99 ½ (±5% ±0.03 ½) 10.0 ½ – 99.9 ½ (±5% ±0.5 ½) 100 ½ – 999 ½ (±5% ±5 ½) 1.00 k½ – 2.00 k½ (±5% ±30 ½) |
Hi Current Line/Earth Loop Test (2 Wire, LRCD220 Only) | Yes |
Supply | 100 V – 480 V |
Nominal Test Current | 15 mA to 2.4 A |
Loop Accuracy | 0.01½ – 9.99 ½ (±5% ±0.03 ½) 10.0½ – 99.9 ½ (±5% ±0.5 ½) 100½ – 999 ½ (±5% ±5 ½) 1.0 k½ – 2.00 k½ (±5% ±30 ½) |
Line/Line (LRCD220 Only) Supply | 100 V – 480 V 45Hz to 65Hz |
Prospective Fault Current (PSCC) | Prospective fault current = Nominal supply voltage/Loop resistance Accuracy is derived from loop test 1 A – 199 A 1 A resolution 0.20 kA – 1.99 kA 10 A resolution 2.0 kA – 19.9 kA 100 A resolution |
RCD Testing | RCD Test Ranges (to EN61557-6) LRCD200/210 10mA, 30mA,100mA, 300mA, 500mA LRCD220 10mA, 30mA,100mA, 300mA, 500mA, 1000mA |
Supply Voltage | LRCD200/210 = 200 V – 280 V 45Hz to 65Hz LRCD220 = 100 V – 280 V 45Hz to 65Hz |