• Store Locator
  • Schedule a
    Demo
  • Add products to your Toolbox to save it for requesting a quote while you keep browsing for other products to complete your list.

    My
    Toolbox
  • PCE Instruments TTC 30 Thermocouple Calibrator 30V DC

    # TTC 30

    • The thermocouple calibrator PCE-TTC 30 can be used to simulate different thermocouples as well as mV. Thus, the thermocouple calibrator can be used for many thermometers and temperature displays with thermocouple connection.
    • The PCE-TTC 30 thermocouple calibrator simulates the most popular thermocouples, including K, T, S. The simulator for thermocouples has an accuracy of 0.3°C / 0.54°F, making it easy to accurately calibrate thermometers.
    • In addition to direct specification of temperatures or extra-low voltages up to 250 mV, the thermocouple calibrator can be operated in ramp mode. Here, the user can decide whether the ramp should be run indefinitely or at defined intervals.
    • The thermocouple calibrator PCE-TTC 30 can be used to simulate different thermocouples as well as mV. Thus, the thermocouple calibrator can be used for many thermometers and temperature displays with thermocouple connection. The PCE-TTC 30 thermocouple calibrator simulates the most popular thermocouples, including K, T, S. The simulator for thermocouples has an accuracy of 0.3°C / 0.54°F, making it easy to accurately calibrate thermometers.
    • In addition to direct specification of temperatures or extra-low voltages up to 250 mV, the thermocouple calibrator can be operated in ramp mode. Here, the user can decide whether the ramp should be run indefinitely or at defined intervals. The miniature USB interface can be used to load the thermocouple calibrator. Furthermore, recorded data can be transmitted via this interface to a PC.
    • Simulation and measurement mode
    • Battery operation
    • For thermocouples E, J, K, T, B, R, S, N, mV
    • 24 V supply for current loops
    • HART compatible
    • Manual mode & ramp function
    • Continuity test
    • Data logger function
    Voltage DC V
    Measuring rangeResolutionAccuracy
    0 … 30V0.001V± 0.02% of reading + 2 Dgt
    Current DC mA
    Measuring rangeResolutionAccuracy
    0 … 24-mA0.001-mA± 0.02% of reading + 2 Dgt
    Measurement / Simulation
    Thermocouples
    Measuring range
    Resolution
    Accuracy
    E
    -200 … 1200°C / -328 … 2192°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    J
    -200 … 1200°C / -328 … 2192°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    K
    -200 … 1372°C / -328 … 2501°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    T
    -200 … 400°C / -328 … 752°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    B
    450 … 1800°C / 842 … 3272°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    R
    0 … 1750°C / 32 … 3182°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    S
    0 … 1750°C / 32 … 3182°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    N
    -200 … 1300°C / -328 … 2372°F
    0.1°C / 0.18°F
    ± 0.3°C / 0.54°F
    mV
    10 … 80 mV
    0.001 mV
    ± 0.02% of reading + 2μV
    -10 … 250 mV
    0.01 mV
    ± 0.02% of reading + 0.02 mV
    Note: The specifications refer to the temperature standard ITS-90
    General specifications
    Display modesMeasurement: mA / V / thermocouple / mV
    Simulation: thermocouple / mV
    Temperature units°C / °F / K
    Cold junction compensation error≤ ± 0.5°C / 0.9°F
    Maximum input voltage30V DC
    Temperature coefficient<30 ppm
    Input impedance measurementThermocouple / mV / V:> 1 MΩ
    Current measurement: 10 Ω
    Response time<100 ms
    Impedance> 4.7 kΩ with thermocouple / mV
    Refresh rate display10 Hz
    Isolation500V DC
    Data storageInternal memory 150000 readings
    InterfaceUSB 2.0
    Display2.4″ TFT LCD
    240 x 320 pixels
    LED illuminated
    Output voltage current loop24V DC / 24mA
    HART mA loop resistance250 Ω ± 20%
    Special featuresStep and ramp function
    Automatic and manual
    mode √x, x2: For the measuring function
    Continuity testAdjustable threshold up to 100 Ω
    Power supply3.7V / 2300mAh Li-ion battery
    Charging timeabout 5 h
    Power adapterInput: 100 … 240V AC / 50/60 Hz
    Output: 5V / 1 A DC
    Battery lifeApprox. 15 h: Simulation and measurement with low LCD illumination
    Approx. 8 h: Measurement with low LCD illumination
    Dimensions162 x 82 x 40 mm
    WeightAbout 300 g / < 1 lb
    Degree of protectionIP20
    Operating conditionsBattery operation: 0 … 55°C / 32 … 131°F, 30 … 90% RH Main operation: 0 … 45°C / 32 … 113 °F, 30 … 90% RH
    Storage conditions-20 .. 60°C / -4 … 140°F, 30 … 90% rh non-condensing
    Heating timeAbout 15 minutes
    • 1 x Thermocouple Calibrator PCE-TTC 30
    • 2 x Laboratory cables with 2 mm plug
    • 1 x Thermocouple adapter to 2 mm plug
    • 2 x Adapter 2 mm on 4 mm laboratory cable
    • 2 x Alligator clips
    • 1 x Mini USB cable
    • 1 x Power supply 5 V / 1 A
    • 1 x Carrying bag
    • 1 x Software
    • 1 x User manual

    Provide us your contact detailsone of our product experts will contact you to discuss your needs or schedule a demonstration at a time that suits

    Product selected: Model #







      *Demonstrations can only be scheduled within the UAE

      Sign up to stay in the know-how of our latest products, updates, special offers, and more.

      • Subscription Form