# MCA 50
Display modes (Measurement) | mA / V / mV / mA (24V) / Switch test / Temperature thermocouple / Resistance thermometer / Frequency / Pulse |
Display modes (Simulation) | mA / V / mV / mA (2 W) / Resistance / Resistance Thermometer / Temperature Thermocouple / Frequency / Pulse |
Maximum input voltage | 30V DC |
Input impedance Measurement | Thermocouple, mV, V, frequency, pulse: 1 MΩ, mA: 10 MΩ |
Response time | <100 ms |
Load impedance | > 4.7 KΩ for thermocouple / mV / V / pulse / frequency, <750 Ω for mA |
Refresh rate display | 10 readings per second |
Isolation | 500V DC |
Data storage | 150000 measured values maximum |
Interface | USB 2.0 |
Display | 3.2in TFT LCD, 240 x 320 pixels, LED illuminated |
Output voltage Current loop | 24V DC / 24-mA |
HART mA loop Resistance | 250 Ω ± 20% |
Special features | Step and ramp function, Automatic and manual, mode √x, x2: For the measuring function |
Continuity test | Adjustable threshold value up to 100 Ω |
Power supply | Rechargeable battery, 3000-mAh 3.7V |
Charging time | Max. 5 hours |
Power adapter | 100 to 240V AC, 50/60 Hz, output 5V DC at 1 A |
Battery life (> 17 h) | Simulation and measurement with low LCD illumination |
Battery life (> 9 h) | Measurement with low LCD illumination |
Dimensions (L x W x H) | 185.6 x 97.1 x 41.3 mm / 7.3 x 3.8 x 1.6 in |
Weight | <500 g / 1.1 lbs |
Protection class | IP20 |
Operating conditions | 0 to 55°C / 32 to 131°F, 30 to 90% RH (non-condensing) |
Storage conditions | -20 to 60°C / -4 to 140°F, 30 to 90% rh (non-condensing) |
Heating time | 5 minutes |