• Store Locator
  • Schedule a
    Demo
  • Add products to your Toolbox to save it for requesting a quote while you keep browsing for other products to complete your list.

    My
    Toolbox
  • FLUKE 80TK Thermocouple Module (Type K)

    # 80TK

    • Converts a DMM to a thermometer
    • For use with type-K thermocouples in low voltage applications (below 24 V AC, 60 V DC)
    • Measurement range: -58 to 1832°F (-50 to 1000°C) (probe dependent)
    • Converts a DMM to a thermometer
    • For use with type-K thermocouples in low voltage applications (below 24 V AC, 60 V DC)
    • Measurement range: -58 to 1832°F (-50 to 1000°C) (probe dependent)
    Measurement Range-58 to 1832°F (-50 to 1000°C)
    Resolution2000 count display: 0.1°F/C in 200 mV range; 1.0°F/C in 2 V range
    3200 count display: 0.1°F/C in 300 mV range; 1.0°F/C in 3 V range
    4000 count display: 0.1°F/C in 400 mV range; 1.0°F/C in 4 V range
    Ambient Operating Range32 to 122°F (0 to 50°C)
    Storage Temperature-40 to 140°F (-40 to 60°C)
    Humidity0 to 90% (32 to 95°F [0 to 35°C])
    0 to 70% (95 to 122°F [35 to 50°C])
    Basic Accuracy (at 23 ±5°C Calibration)-4 to 662°F (-20 to 350°C): 0.5% ± 3.6°F (0.5% ± 2°C)
    664 to 932°F (351 to 500°C): 1.75% ± 3.6°F (1.75% ± 2°C)
    933 to 1832°F (501 to 1000°C): 2.0% ± 3.6°F (2.0% ± 2°C)
    -58 to -5.8°F (-50 to -21°C): 2.5% ± 3.6°F typical (2.5% ± 2°C typical)
    BatteryType: Standard 9 V Battery (NEDA #1604, 6F22, 006P)
    Life: 1600+ hours, 6.5 V minimum
    Output¾” spaced banana plugs, 10 MΩ minimum load resistance
    Overall Length4 ⅝” (11.75 mm)
    Weight4.5 oz (126 g)