• Store Locator
  • Schedule a
    Demo
  • Add products to your Toolbox to save it for requesting a quote while you keep browsing for other products to complete your list.

    My
    Toolbox
  • FLUKE 712B RTD Process Calibrator;

    # 712B

    • Capable of measuring and simulating 13 different RTD types and resistances.
    • Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal.
    • Remembers power down settings to easily restart tests when powered up

    Fluke 712B RTD Calibrator provides highly accurate, easy-to-use, single-function RTD temperature calibration

    Fluke 712B RTD Calibrator is a handheld, battery-operated instrument that measures and sources a variety of RTD types and resistances. It also includes an isolated channel to measure 4-20 mA whiles sourcing a temperature signal. It offers configurable 0% and 100% source settings for quick 25% linearity checks. It also provides linear ramp and 25% step auto ramp functionality based on 0% and 100% settings.

    Add to that dual inputs, an easy to view backlit display, plus the ability to memorize settings at power down for easy restart make the 712B the ideal test tool for temperature calibration professionals who want a highly accurate, easy-to-use, single function RTD temperature calibrator.

    Other useful features:

    • Dual inputs and backlit display for easy interpretation of measurements
    • Remembers power down settings to easily restart tests when powered up
    • Provides linear ramp and 25% step auto ramp based on 0% and 100% settings
    • Provides one-year and two-year specifications and traceable certification of calibration
    • Highly accurate, full-featured , easy-to-use single function RTD temperature calibrator
    • Measures and simulates 13 different RTD types and resistance
    • Measures 4 to 20 mA signals while simultaneously sourcing a temperature signal
    • Comes with a magnetic hanging tool designed in for convenient hands-free operation
    • Features configurable 0% and 100% source settings for quick 25% linearity checks
    General Specifications
    Maximum voltage applied between any terminal and earth ground or between any two terminals30 V
    Operating temperature-10°C to 50°C
    Storage temperature-30°C to 60°C
    Operating altitude2,000 m
    Storage altitude12,000 m
    Relative humidity (% RH operating without condensation)Non condensing
    90% (10°C to 30°C)
    75% (30°C to 40°C)
    45% (40°C to 50°C)
    (Without condensation)
    Vibration requirementsMIL-T-28800E, Class 2
    Drop text requirements1 m
    IP RatingIEC 60529: IP52
    Electromagnetic environmentIEC 61326-1, Portable
    SafetyIEC 61010-1, Max 30 V to earth, Pollution Degree 2
    Power supply4 AA NEDA 1.5A IEC LR6 batteries
    Size (H x W x L)52.5 x 84 x 188.5 mm
    Weight515 g
    DC mA Measurement
    Resolution0-24 mA
    Range0.001 mA
    Accuracy (% of reading + counts)0.010% + 2 μA
    Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( < 18°C or >28°C)
    Ohms Measurement
    Ohms rangeAccuracy (% of reading +counts)
    0.00 Ω to 400.00 Ω0.015% + 0.05 Ω
    400.0 Ω to 4000.0 Ω0.015% + 0.5 Ω
    Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications.
    Temperature coefficient±(0.002% of reading + 0.002% of range) / °C (< 18°C or > 28°C)
    Ohms Source
    Ohms range1.0 Ω to 400.0 Ω
    1.00 Ω to 400.00 Ω
    400.0 Ω to 1500.0 Ω
    1500.0 Ω to 4000.0 Ω
    Excitation current from measurement device0.1 mA to 0.5 mA
    0.5 mA to 3 mA
    0.05 mA to 0.8 mA
    0.05 mA to 0.4 mA
    Accuracy (% of reading + counts)0.015% + 0.1 Ω
    0.015% + 0.05 Ω
    0.015% + 0.5 Ω
    0.015% + 0.5 Ω
    Resolution0.00 Ω to 400.00 Ω0.01 Ω
    400.0 Ω to 4000.0 Ω0.1 Ω
    Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( <18°C or >28°C)
    Supports pulsed transmitters and PLCs with pulse times as short as 5 ms
    RTD Input and Output
    RTD Type (α)Range (°C)Measure (°C)Source (°C)
    1-year2-yearSource current1-year2-year
    10 Ω Pt (385)-200 to 100°C1.5°C3°C1 mA1.5°C3°C
    100 to 800°C1.8°C3.6°C1 mA1.8°C3.6°C
    50 Ω Pt (385)-200 to 100°C0.4°C0.7°C1 mA0.4°C0.7°C
    100 to 800°C0.5°C0.8°C1 mA0.5°C0.8°C
    100 Ω Pt (385)-200 to 100°C0.2°C0.4°C1 mA0.2°C0.4°C
    100 to 800°C0.015%
    +0.18°C
    0.03%
    +0.36°C
    1 mA0.015%
    +0.18°C
    0.03%
    +0.36°C
    200 Ω Pt (385)-200 to 100°C0.2°C0.4°C500 μA0.2°C0.4°C
    100 to 630°C0.015%
    +0.18°C
    0.03%
    +0.36°C
    500 μA0.015%
    +0.18°C
    0.03%
    +0.36°C
    500 Ω Pt (385)-200 to 100°C0.3°C0.6°C250 μA0.3°C0.6°C
    100 to 630°C0.015%
    +0.28°C
    0.03%
    +0.56°C
    250 μA0.015%
    +0.28°C
    0.03%
    +0.56°C
    1000 Ω Pt (385)-200 to 100°C0.2°C0.4°C250 μA0.2°C0.4°C
    100 to 630°C0.015%
    +0.18°C
    0.03%
    +0.36°C
    250 μA0.015%
    +0.18°C
    0.03%
    +0.36°C
    100 Ω Pt (3916)-200 to 100°C0.2°C0.4°C1 mA0.2°C0.4°C
    100 to 630°C0.015%
    +0.18°C
    0.03%
    +0.36°C
    1 mA0.015%
    +0.18°C
    0.03%
    +0.36°C
    100 Ω Pt (3926)-200 to 100°C0.2°C0.4°C1 mA0.2°C0.4°C
    100 to 630°C0.015%
    +0.18°C
    0.03%
    +0.36°C
    1 mA0.015%
    +0.18°C
    0.03%
    +0.36°C
    10 Ω Cu (427)-100 to 260°C1.5°C3°C1 mA1.5°C3°C
    120 Ω Ni (672)-80 to 260°C0.15°C0.3°C1 mA0.15°C0.3°C
    50 Ω Cu (427)-180 to 200°C0.4°C0.7°C1 mA0.4°C0.7°C
    100 Ω Cu (427)-180 to 200°C0.2°C0.4°C1 mA0.2°C0.4°C
    YSI40015 to 50°C0.2°C0.4°C250 μA0.2°C0.4°C
    1. Sensor inaccuracies not included.
    2. Resolution: 0.1°C.
    3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0°C (Pt10 and Cu10), 0.6°C (Pt50 and Cu50), 0.4°C (Other RTD types) to the specifications.
    4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range).
    5. Temperature Coefficient: ±0.05°C /°C for measure, ±0.05°C /°C (< 18°C or > 28°C) for source.
    6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms.
    • Magnetic hanging tool
    • Batteries
    • Manual
    • Traceable calibration certificate
    • Test Lead

    Provide us your contact detailsone of our product experts will contact you to discuss your needs or schedule a demonstration at a time that suits

    Product selected: Model #







      *Demonstrations can only be scheduled within the UAE

      Sign up to stay in the know-how of our latest products, updates, special offers, and more.

      • Subscription Form